Adult Pelvis with Organs for X-Ray CT, MRI
The Adult Pelvis with Organs Phantom is designed based on an average adult human (male). The synthetic bones used in this phantom are made out of our patented material, which is best suited for the X-Ray/CT, and MRI imaging applications. This product can be further customized to match any project’s specific requirements.
The Adult Pelvis with Organs Phantom is designed based on an average adult human (male). It is made out of realistic patented bone material that is suitable for X-RAY CT and MRI imaging applications. It can be used as a tool to conduct medical imaging research, radiology training, testing, and calibration purposes. This product can be further customized to match our client’s specific requirements.
In terms of MRI applications, the phantom tissues have realistic T2 relaxation time values, which makes this product the best fit for any T2-weighted MRI imaging methods. Very good results can also be achieved with proton-density imaging methods. The phantom can still be imaged with T1-weighted methods, but the T1 values are less realistic, and they are within the range of about 100 ms.
The bones will consist of a realistic three-layered structure with inner porosity, which can be adjusted according to the requirement of the particular project. The phantom can also be customized in size and shape (i.e. based on any individual’s CT scan).
Note: By default, the phantom comes in a translucent amber appearance. The skin feature can be included upon request.
| Type of Tissue | Sound Velocity [m/s] | Density [g/cm³] | Hardness [Shore 00] | T2 [ms] | Hounsfield Value | Attenuation Measured at 2.25 MHz [dB/cm] |
| Cortical Bone | 3000 ± 30 | 2.31 | N/A | N/A | 1300 | 6.4 ± 0.3 |
| Trabecular Bone | 2800 ± 50 | 2.03 | N/A | N/A | 1300 | 21 ± 2 |
| Soft Tissue | 1410 ± 10 | 1.02 | 60 | 50 ms | 25 | 1.7 ± 0.2 |
| Thermal Properties of Bone-Mimicking Material | |||||
| Thermal Conductivity | Volumetric Specific Heat Capacity | Thermal Diffusivity | Thermal Resistivity | Specific Heat | Speed of Sound |
| 0.776 W/ m K | 1.040 MJ/ m^3 K | 0.746 mm^2/ s | 1.289 m K/ W | 0.978 J/ g Deg Celsius | 3070 m/s |
| HU Values of the Tissue-Mimicking Materials | ||
| S.No. | Tissue Type | HU Value (average) |
| 1 | Trabecular Bone | 800 |
| 2 | Cortical Bone | 1300 |
Sample scans of this phantom. Full scan galleries available on request.